Related Experiment Video
Updated: Dec 30, 2025

Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
Drug metabolite identification using ultrahigh-performance liquid chromatography-ultraviolet spectroscopy and
Stephen D Wilkinson1, Scott Martin1, Alexandra L Orton1
1DMPK/ADME Research and Early Development, Oncology R&D, AstraZeneca, Cambridge, UK.
Rationale:
To capture all metabolites in metabolite identification studies, MS/MS information is required in both positive and negative ionization mode, usually involving several sample injections to gain all information about samples. A high-resolution and high mass accuracy quadrupole/linear trap/Orbitrap tribrid instrument was used to gain this information in a novel single injection 'capture-all' approach to metabolite identification.
Methods:
Diclofenac, a model compound, was incubated in human and rat hepatocytes. These incubated samples were run using an ultrahigh-performance liquid chromatography/ultraviolet (UHPLC-UV) system coupled to a Thermo Fusion tribrid mass spectrometer. Five parallel scans were used: positive and negative ion full scan, data-dependent MS/MS, both high energy dissociation and collision-induced dissociation, and data-independent all ion fragmentation (AIF) spectra were collected in positive and negative ion mode.
Results:
Nine metabolites were identified; a metabolite observed in the UV trace, but not positive ion full scan MS, was detected in the same sample injection by negative ion full scan MS. This was identified as a sulphate metabolite, and the corresponding negative ion AIF allowed for some structural elucidation. The use of a photo-diode array (PDA) detector allowed for spectral assessment in case of changes in absorbance spectra, and the subsequent semi-quantification of metabolites.
Conclusions:
This method provided good-quality MS/MS data across the m/z range in both positive and negative ion mode. The addition of both negative ion full scan MS and negative ion MS/MS allowed for the detection and structural elucidation of metabolites not observed in positive ion mode. The use of the PDA detector allowed for the semi-quantification of metabolites.
More Related Videos
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Tandem Mass Spectrometry
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
High-Resolution Mass Spectrometry (HRMS)
High-Performance Liquid Chromatography: Types of Detectors