An implantable semiconductor beta-radiation detector.
The American Journal of Physiology
|January 1, 1977
Summary
This study introduces an implantable beta-radiation detector for measuring regional blood flow in animals. The device shows good agreement with mechanical systems but has limitations in high-flow conditions and myocardial blood flow comparisons.
Area of Science:
- Biomedical Engineering
- Medical Physics
- Physiology
Background:
- Accurate measurement of regional blood flow is crucial for understanding physiological processes and disease.
- Existing methods for blood flow measurement have limitations in accuracy, invasiveness, or applicability to specific tissues.
Purpose of the Study:
- To describe an implantable beta-radiation detector for regional blood flow measurement in experimental animals.
- To evaluate the accuracy and reliability of this novel detector.
- To compare its performance with established blood flow assessment techniques.
Main Methods:
- Development and implantation of a lithium-drifted silicon diode detector.
- Use of a beta-emitting isotope (e.g., 85Kr) for indicator clearance principle.
- Calibration against a mechanical system and in vivo comparisons with precordial counter and microsphere techniques.
Main Results:
- The detector demonstrated excellent agreement with mechanical calibration up to 600 ml/100 g per min.
- At very high flow rates, flow was underestimated by a maximum of 10%.
- In vivo comparisons of myocardial blood flow showed poor agreement with precordial counter and microsphere methods.
Conclusions:
- The implantable beta-radiation detector is a viable tool for regional blood flow measurement in experimental animals.
- Discrepancies in myocardial blood flow measurements may be due to tissue heterogeneity or methodological limitations.
- The device maintained functionality for over 50 days in chronic animal studies.