Related Experiment Video
Updated: Dec 30, 2025

Integrated Compensatory Responses in a Human Model of Hemorrhage
Published on: November 20, 2016
Validating clinical threshold values for a dashboard view of the compensatory reserve measurement for hemorrhage
Victor A Convertino1, Mallory R Wampler, Michael Johnson
1From the Division of Trauma and Emergency Surgery (M.R.W., M.J., S.N., J.G.M., B.J.E.), UT Health, San Antonio, Texas; Department of Vascular Surgery (A.A.), Cleveland Clinic, Cleveland, Ohio; Battlefield Health and Trauma Center for Human Integrative Physiology (V.A.C., T.D.L., K.R.S., C.C.), US Army Institute of Surgical Research, San Antonio, Texas; and Department of Medicine and Surgery (K.K.C.), Uniformed Services University, Bethesda, Maryland.
Background:
Compensatory reserve measurement (CRM) is a novel noninvasive monitoring technology designed to assess physiologic reserve using feature interrogation of arterial pulse waveforms. This study was conducted to validate clinically relevant CRM values with a simplified color-coded dashboard view.
Methods:
We performed a prospective observational study of 300 injured patients admitted to a level I trauma center. Compensatory reserve measurement was recorded upon emergency department admission. Data collected to complement the analysis included patient demographics, vital signs, lifesaving interventions, Injury Severity Score (ISS), and outcomes. Threshold values of CRM were analyzed for predictive capability of hemorrhage.
Results:
A total of 285 patients met the inclusion criteria. Mean age of the population was 47 years, and 67% were male. Hemorrhage was present in 32 (11%), and lifesaving intervention was performed in 40 (14%) patients. Transfusion of packed red blood cells was administered in 33 (11.6%) patients, and 21 (7.4%) were taken to the operating room for surgical or endovascular control of hemorrhage. Statistical analyses were performed to identify optimal threshold values for three zones of CRM to predict hemorrhage. Optimal levels for red, yellow, and green areas of the dashboard view were stratified as follows: red if CRM was less than 30%, yellow if CRM was 30% to 59%, and green if CRM was 60% or greater. Odds of hemorrhage increased by 12-fold (odds ratio, 12.2; 95% confidence interval, 3.8-38.9) with CRM less than 30% (red) and 6.5-fold (odds ratio, 6.5; 95% confidence interval, 2.7-15.9) with CRM of equal to 30% to 59% (yellow) when compared with patients with CRM of 60% or greater. The area under the receiver operating characteristic curve for three-zone CRM was similar to that of continuous CRM (0.77 vs. 0.79) but further increased the ability to predict hemorrhage after adjusting for ISS (area under the receiver operating characteristic curve, 0.87).
Conclusion:
A three-zone CRM could be a potentially useful predictor of hemorrhage in trauma patients with added capabilities of continuous monitoring and a real-time ISS assessment. These data substantiate easily interpretable threshold dashboard values for triage with potential to improve injury outcomes.
Level Of Evidence:
Diagnostic, level II.
Related Concept Videos
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.