Reflotron cholesterol measurement evaluated as a screening technique
F A Sedor1, C M Holleman, S Heyden
1Hospital Laboratories, Duke University Medical Center, Durham, NC 27710.
Clinical Chemistry
|December 1, 1988
Summary
The Reflotron analyzer demonstrates reliable cholesterol measurement performance across various sample types and personnel training levels. Analytical precision and comparability with established methods support its clinical utility for cholesterol testing.
Area of Science:
- Clinical Chemistry
- Analytical Performance Evaluation
- Diagnostic Instrumentation
Background:
- Accurate cholesterol measurement is crucial for cardiovascular disease risk assessment.
- Point-of-care testing devices require rigorous analytical validation.
- The Boehringer Mannheim Diagnostics Reflotron analyzer is a widely used diagnostic instrument.
Purpose of the Study:
- To evaluate the analytical performance of the Reflotron analyzer for cholesterol determination.
- To assess the impact of different sample types and personnel training on measurement accuracy.
- To identify potential interferents affecting Reflotron cholesterol assay results.
Main Methods:
- Assessed within-day and between-day coefficients of variation (CVs) for whole blood, serum, and plasma.
- Evaluated performance using commercial quality-control materials.
- Investigated interference from heparin, hematocrit, bilirubin, hemoglobin, and triglycerides.
- Compared Reflotron results against the Ektachem procedure using samples analyzed by trained personnel and laymen.
Main Results:
- Achieved low coefficients of variation (CVs) for cholesterol measurements in whole blood, serum, and plasma (1.4%–2.6%).
- Demonstrated good comparability (r = 0.950–0.980) with the Ektachem method when analyzed by trained personnel and students.
- Observed decreased performance (r = 0.880) when both sample collection and analysis were performed by untrained laymen.
Conclusions:
- The Reflotron analyzer exhibits acceptable analytical performance for cholesterol measurement.
- Measurement accuracy is maintained with trained personnel, supporting its use in clinical settings.
- Careful sample handling and trained operators are essential for optimal Reflotron assay reliability.