Related Experiment Video
Updated: Dec 30, 2025

Author Spotlight: Optimizing EAS with Long Electrodes for Enhanced Cochlear Coverage and Hearing Preservation
Published on: October 11, 2024
Clinical Experience of Vibroplasty With Direct Coupling to the Oval Window Without Use of a Coupler
Sang Hyun Kwak1, Young Min Moon1, Gi-Sung Nam1
1Department of Otorhinolaryngology, Yonsei University College of Medicine, Seoul, South Korea.
Objectives/Hypothesis:
To investigate the efficacy of direct implantation of a Vibrant Soundbridge (VSB) implant in the oval window (OW) without the use of an OW coupler in patients with severe mixed hearing loss.
Study Design:
Retrospective chart review METHODS: A total of 62 patients underwent VSB implantation between July 2016 and December 2018 at Severance Hospital in Seoul, South Korea. Among them, eight patients (nine ears) with moderate-to-severe mixed hearing loss were implanted with a VSB directly in the OW. A floating mass transducer (FMT) was attached to the stapes footplate and covered with tragal cartilage. The outcomes were evaluated using pure-tone audiogram and speech audiogram preoperatively and postoperatively. Word recognition score (WRS; % correct) were measured at the most comfortable loudness (MCL) level to evaluate speech perception.
Results:
All cases posed difficulty with round window vibroplasty during surgery, and eventually, an FMT was appropriately placed in the OW without a coupler. Preoperative and postoperative bone conduction thresholds were not different. VSB-aided threshold improved in terms of functional and effective gains. Interestingly, four cases showed improved air conduction thresholds without the use of a VSB. In addition, MCL level with a VSB was significantly lower than that with a hearing aid, and VSB-aided WRS improved over time.
Conclusions:
Direct implantation of a VSB in the OW without the use of a coupler showed favorable hearing outcomes, and the OW vibroplasty was safe. Direct OW vibroplasty without a coupler is a reliable procedure and can be a good option for hearing rehabilitation in patients with severe mixed hearing loss.
Level Of Evidence:
4 Laryngoscope, 2020.
More Related Videos
09:07Author Spotlight: Advancing Endoscopic Ossiculoplasty – Techniques, Innovations, and Practical Guidance for Clinical Integration
Published on: January 26, 2024
09:06Cochlear Implant Surgery and Electrically-evoked Auditory Brainstem Response Recordings in C57BL/6 Mice
Published on: January 9, 2019
Related Concept Videos
The Auditory Ossicles
The aptly named stapes look very much like a stirrup. The three ossicles are unique to mammals, and each plays a role in...
Patch Clamp
In this method, a glass micropipette containing electrolyte solution is tightly sealed against a small portion of the cell membrane. As a result, a patch of the cell...
Anatomy of the Ear
The Cochlea