Related Experiment Video
Updated: Dec 30, 2025

Author Spotlight: Advancements in Multiplex Detection of Respiratory Viruses
Published on: November 10, 2023
Development and evaluation of a panel of multiplex one-tube nested real time PCR assay for simultaneous detection of
Li Zhao1, Gui-Xia Li2, Ji Wang3
1Department of Blood Transfusion, Children's Hospital of Hebei Province, Shijiazhuang, Hebei, China.
Abstract:
Multiplex real-time quantitative polymerase chain reaction (mRT-qPCR) assay is commonly used to detect respiratory viruses, however, the sensitivity is limited for most reports. A panel of locked nucleic acid based multiplex closed one-tube nested real-time PCR (mOTNRT-PCR) assay consisting of five separate internally controlled RT-qPCR assays was developed for detection of 14 respiratory viruses. The sensitivity and reproducibility of mOTNRT-PCR panel were evaluated using plasmid standards and the specificity was evaluated using clinical samples. The clinical performance of mOTNRT-PCR panel was further evaluated with 468 samples collected from patients with an acute respiratory infection and compared with individual real-time PCR (RT-qPCR) assay. The analytical sensitivities of mOTNRT-PCR panel ranged from 2 to 20 copies/reaction, and no cross-reaction with common respiratory viruses was observed. The coefficients of variation of intra-assay and inter-assay were between 0.35% and 8.29%. Totally 35 clinical samples detected by mOTNRT-PCR assay panel were missed by RT-qPCR and confirmed true positive by sequencing of nested PCR products. The mOTNRT-PCR assay panel provides a more sensitive and high-throughput method for the detection of 14 respiratory viruses.