Related Experiment Video
Updated: Dec 30, 2025

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
A fast method for the detection of irinotecan in plasma samples by combining solid phase extraction and differential
Gregorio Bonazza1, Stefano Tartaggia2, Giuseppe Toffoli2
1Department of Molecular Science and Nanosystems, Ca' Foscari University of Venice, Via Torino 155, Mestre-Venezia, 30172, Venice, Italy.
A new, fast method accurately detects irinotecan (CPT-11) in plasma using solid phase extraction and electrochemical detection. This technique is suitable for clinical applications in cancer chemotherapy monitoring.
Area of Science:
- Analytical Chemistry
- Clinical Chemistry
- Pharmacology
Background:
- Irinotecan (CPT-11) is a vital chemotherapy drug for solid tumors.
- Accurate monitoring of CPT-11 plasma levels is crucial for effective cancer treatment.
- Existing detection methods may lack speed or specificity for clinical settings.
Purpose of the Study:
- To develop and validate a rapid and reliable method for quantifying irinotecan (CPT-11) in human plasma.
- To assess the method's suitability for monitoring CPT-11 levels during chemotherapy.
- To ensure the method effectively removes interfering substances commonly co-administered with CPT-11.
Main Methods:
- Solid phase extraction (SPE) using acetonitrile as eluent for CPT-11 isolation from plasma.
- A cleaning step was incorporated to remove endogenous compounds and co-therapeutics (5-FU, FA).
- Electrochemical detection via differential pulse voltammetry at a glassy carbon electrode (GCE) in a basic acetonitrile medium.
Main Results:
- A well-defined oxidation peak for CPT-11 was achieved, free from interference from metabolites.
- The method demonstrated good linearity with a limit of detection (LOD) of 1.10 × 10⁻⁷ M and a limit of quantification (LOQ) of 3.74 × 10⁻⁷ M.
- Successful determination of CPT-11 concentrations in plasma samples from an oncology patient at 30 and 180 minutes post-infusion.
Conclusions:
- The developed SPE-electrochemical detection method offers a fast and accurate approach for CPT-11 analysis in plasma.
- This method shows significant potential for routine clinical application in cancer patient management.
- The technique effectively addresses challenges of interference from co-administered drugs and endogenous compounds.
Related Concept Videos
Ion-Exchange Chromatography
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...