Abstract:
Regional cerebral oxygenation index (rSO2) based on near-infrared spectroscopy (NIRS) is frequently used to detect low venous oxyhemoglobin saturation (ScvO2). We compared the performance of 2 generations of NIRS devices. Clinically obtained, time-matched cerebral rSO2 and ScvO2 values were compared in infants monitored with the FORE-SIGHT (n = 73) or FORE-SIGHT ELITE (n = 47) by linear regression and Bland-Altman analyses. In both devices, cerebral rSO2 correlated poorly with measured ScvO2 (FORE-SIGHT partial correlation 0.50 [95% confidence interval {CI}, 0.40-0.58]; FORE-SIGHT ELITE partial correlation 0.47 [0.39-0.55]) and mean bias was +8 (standard deviation [SD] 13.2) for FORE-SIGHT and +14 (SD 12.5) for FORE-SIGHT ELITE. When ScvO2 was <30%, rSO2 was <40 in 8% of FORE-SIGHT ELITE readings. Future NIRS should be validated in more hypoxic cohorts.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
09:04Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...