Respiratory Volume Monitoring in the Perioperative Setting Across Multiple Centers
Chunyuan Qiu1, Eugene Cheng2, Shawn R Winnick3
1Department of Anesthesiology, Kaiser Permanente Baldwin Park Medical Center, Baldwin Park, California. chunyuan.x.qiu@kp.org.
Background:
The prevalence of nuisance (technical) alarms is the leading cause of alarm fatigue resulting in decreased awareness and a reduction in effective care. The Joint Commission identified in their National Patient Safety goals alarm fatigue as a major safety issue. The introduction of noninvasive respiratory volume monitoring (RVM) has implications for effective perioperative respiratory status management. We evaluated this within the Kaiser Permanente health system.
Methods:
This observational study was conducted at 4 hospitals in the Kaiser Permanente system. Standard data from RVM, pulse oximetry, and capnography were collected postoperatively in the post-anesthesia care unit (PACU) and/or on the general hospital floor. Device-specific alarm types, rates, and respective actions were recorded and analyzed by non-study staff.
Results:
RVM was applied to 247 subjects (143 females, body mass index 32.3 ± 8.7 kg/m2, age 60.9 ± 13.9 y) providing 2,321 h. RVM alarms occurred 605 times (0.25 alarms/h); 64% were actionable and addressed, 17% were not addressed, 13% were self-resolved, and only 6% were nuisance. In a subgroup, RVM completed all 127 h of monitoring, whereas oximetry with capnography only completed 51 h with 12.9 alarms/h (73% nuisance). The overall RVM alarm rate was significantly lower than with either pulse oximeters or capnography monitors. We saw a nearly 1,000-fold reduction in nuisance alarms compared to capnography and a 20-50-fold reduction in nuisance alarms compared to pulse oximetry.
Conclusions:
Our study indicates that alarm fatigue due to nuisance alarms continues to be a clinical challenge in perioperative settings. Among the 3 common technologies for respiratory function monitoring, RVM had the lowest rate of overall technical alarms and the highest rate of compliance. Furthermore, with early interventions, none of the subjects monitored with RVM suffered any negative outcomes.
More Related Videos
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Respiratory Volumes and Capacities
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Volumes and Capacities I
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...