In vitro performance of combinations of anti-siphon devices with differential pressure valves in relation to the
I Fiss1, M Vanderheyden2, C von der Brelie2
1Department of Neurosurgery, Universitätsmedizin Göttingen, Georg-August University Göttingen, 37099, Göttingen, Germany. ifweb@web.de.
Background:
Programmable differential pressure (DP) valves combined with an anti-siphon device (ASD) represent the current standard of care in preemtping overdrainage associated with ventriculoperitoneal shunting for hydrocephalus.
Objective:
We aimed to provide comparative data of four combinations of two ASDs of different working principles in combination with two DP valves in an in vitro model in order to achieve a better understanding of the flow characteristics and potential clinical application.
Methods:
We analyzed the flow performance of four possible combinations of two DP valves (CHPV [HM]; proGAV 2.0[PG]) in combination with either a gravity-regulated (Shuntassistant [SA]) or a flow-regulated (SiphonGuard [SG]) ASD in an in vitro setup. A DP between 4 and 60 cmH2O was generated, and the specific flow characteristics were measured. In addition, the two combinations with gravity-regulated ASDs were measured in defined spatial positions.
Results:
Flow characteristics of the SA combinations corresponded to the DP in linear fashion and to the spatial position. Flow characteristics of the SG combinations were dependent upon the DP in a non-linear fashion and independent of the spatial position. Highest mean flow rate of the PG-SG- (HM-SG-) combination was 1.41 ± 0.24 ml/min (1.16 ± 0.06 ml/min). The mean flow rates sharply decreased with increasing inflow pressure and subsequently increased slowly up to 0.82 ± 0.26 ml/min (0.77 ± 0.08 ml/min).
Conclusion:
All tested device combinations were able to control hydrostatic effect and prevent consecutive excessive flow, to varying degrees. However, significant differences in flow characteristics can be seen, which might be relevant for their clinical application.
Related Concept Videos
Pipe Flowrate Measurement
The orifice meter is a simple,...
Bernoulli's Equation: Problem Solving
The first step is to compute the cross-sectional areas of the pipe and the Venturi throat to analyze the pressure difference indicated by the pressure gauge. Next, the continuity equation is...
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
Design Example: Forces in Sluice Gate
Key variables in...
Design Example: Designing a Residential Plumbing System
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...