Measuring the Complete-arch Distortion of an Optical Dental Impression
Published on: May 30, 2019
Comparison of the 3D triple-scan protocol and the impression replica technique for 3-unit tooth-supported fixed
Per Svanborg1, Miranda Andersson1, Therese Reinedahl1
1Department of Prosthodontics/Dental Materials Science, Institute of Odontology, the Sahlgrenska Academy, University of Gothenburg, Göteborg, Sweden.
Abstract:
Objective: The aim of this study was to compare the fit measurements from the 3 D triple-scan protocol and the impression replica technique, by measuring the fit of three-unit tooth-supported fixed dental prostheses. Materials and methods: The results from an earlier study using the triple-scan protocol were compared to results from the impression replica technique for absolute marginal gap, cervical area gap and internal gap. Results: There were no differences for cervical area gap or internal gap, but for absolute marginal gap the impression replica technique obtained significantly smaller fit values. Conclusion: The impression replica technique and the triple-scan protocol may both be used to measure the fit of tooth-supported restorations. However, the scanner used for the triple-scan protocol must be able to obtain scan points at the outermost edge of the restorations when used for absolute marginal gap.