Related Experiment Video
Updated: Dec 29, 2025

Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Bicuspid aortic valve sizing for transcatheter aortic valve implantation: Development and validation of an algorithm
Anna S Petronio1, Marco Angelillis1, Ole De Backer2
1Cardiothoracic and Vascular Department, University Hospital Pisa, Italy.
Background:
No indication are available for transcatheter aortic valve implantation (TAVI) sizing in bicuspid aortic valve (BAV). Aim of the study is to develop and validate a Multi-Slice Computed Tomography (MSCT)-based algorithm for transcatheter heart valve (THV) sizing in patients with stenotic BAV under evaluation for TAVI.
Methods:
A two steps method was applied: 1)evaluation of a cohort of 19 consecutive patients with type I BAV stenosis undergoing TAVI through pre and post-procedural MSCT, and development of an algorithm for THV sizing; 2)validation of the algorithm on a new cohort of 21 patients.
Results:
In the first cohort, a high correlation was found between the raphe-level area measured at pre-procedural MSCT and the smallest THV area measured at post-procedural MSCT (p < 0.001). Moreover, reduced THV expansion was observed among patients with higher calcium burden (p = 0.048). Then, a new algorithm for TAVI sizing in BAV was develop (CASPER: Calcium Algorithm Sizing for bicusPid Evaluation with Raphe). This algorithm is based on the reassessment of the perimeter/area derived annulus diameter, according to three main anatomical features: 1) the ratio between raphe length and annulus diameter; 2)calcium burden; 3)calcium distribution in relation to the raphe. The algorithm was then validated in a new cohort of 21 patients, achieving 100% of procedural success and excellent TAVI performance.
Conclusion:
MSCT assessment of raphe length, calcium burden and its distribution is of crucial relevance in the pre-procedural evaluation of patients with BAV. These anatomical features can be combined in a new and simple algorithm for TAVI sizing.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
06:04Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025