Related Experiment Video
Updated: Dec 29, 2025

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Clinical outcomes of Watchman vs. Amplatzer occluders for left atrial appendage closure (WATCH at LAAC)
Caroline Kleinecke1, Jiangtao Yu1, Philip Neef2,3
1Department of Cardiology, REGIOMED Klinikum Lichtenfels, Lichtenfels, Germany.
Aims:
This study compares clinical outcomes of Watchman vs. Amplatzer devices for left atrial appendage closure (LAAC).
Methods And Results:
Of two real-world registries, the Watchman registry Lichtenfels, Germany, and the Amplatzer registry Bern-Zurich, Switzerland, 303 and 333 consecutive patients, respectively, were included. After a 1:1 propensity score matching, 266 vs. 266 patients were compared by use of the predefined primary efficacy endpoint of stroke, systemic embolism and cardiovascular/unexplained death, the primary safety endpoint of major peri-procedural complications and major bleeding events at follow-up, and the combined hazard endpoint, a composite of all above-mentioned hazards. Mean age was 75.3 ± 7.8 (Watchman) vs. 75.1 ± 9.9 (Amplatzer) years, CHA2DS2-VASc score 4.5 ± 1.7 vs. 4.5 ± 1.5, and HAS-BLED score 3.2 ± 1.0 vs. 3.2 ± 1.0. At a mean follow-up of 2.4 ± 1.3 vs. 2.5 ± 1.5 years and 1.322 patient-years, the primary endpoints of efficacy [40/646, 6.2% [Watchman] vs. 43/676, 6.4% [Amplatzer]; hazard ratio (HR), 1.02; 95% confidence interval (CI), 0.66-1.58; P = 0.92] and safety (33/646, 5.1% vs. 30/676, 4.4%; HR, 0.57; 95% CI, 0.29-1.11; P = 0.10), as well as the combined hazard endpoint (69/646, 10.7% vs. 66/676, 9.8%; HR, 0.80; 95% CI, 0.55-1.12; P = 0.26) were similar for both groups.
Conclusion:
This study suggests comparable efficacy and safety of the Watchman and Amplatzer devices.
Related Concept Videos
Mitral Stenosis III: Medical Management
Aortic Regurgitation III: Medical Management