Emergency department non-invasive cardiac output study (EDNICO): an accuracy study
David McGregor1, Shrey Sharma2, Saksham Gupta2
1Queen Mary University London and Barts Health NHS Trust, London, UK. david.mcgregor@barsthealth.nhs.uk.
Non-invasive cardiac output monitoring methods in the emergency department (ED) showed low accuracy in detecting fluid responders. Further research is needed to validate these tools for improving patient outcomes in critical conditions like sepsis.
Area of Science:
- Emergency Medicine
- Critical Care
- Cardiovascular Physiology
Background:
- Limited data exists on non-invasive cardiac output monitoring in the emergency department (ED).
- Assessing fluid responsiveness is crucial for managing critically ill patients.
Purpose of the Study:
- To evaluate the accuracy of five non-invasive methods for detecting fluid responsiveness in ED patients.
- To compare these methods against a reference standard (echocardiography-derived velocity time integral).
Main Methods:
- Prospective observational study in non-ventilated adult ED patients requiring fluid resuscitation.
- Five non-invasive methods assessed: carotid artery blood flow, aortic Doppler, bioreactance, plethysmography, and IVC collapsibility.
- Sensitivity and specificity determined against the reference standard.
Main Results:
- None of the five non-invasive methods reliably identified fluid responders.
- Sensitivity ranged from 41.2% to 63.2%, and specificity from 45% to 63.6%.
- Low agreement (Cohen's Kappa) was observed across all methods.
Conclusions:
- Current non-invasive methods lack the reliability to accurately identify fluid responders in the ED.
- Development of accurate non-invasive tools is essential for optimizing fluid management in conditions like sepsis.
- Results highlight the need for further validation studies of these monitoring techniques.
More Related Videos
Related Concept Videos
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Acute Coronary Syndrome III: Diagnostic Studies
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...