Effect of Manual Upper Eyelid Elevation on Intraocular Pressure Measurement by Four Different Tonometers
Shunsuke Nakakura, Etsuko Mori1, Yuki Fujio1
1Department of Ophthalmology, Saneikai Tsukazaki Hospital, Himeji, Japan.
Significance:
This study is the first to show that the manual upper eyelid elevation (manual UEE) that is commonly used to prevent disruption of the IOP measurement due to blinking or upper eyelid contact with the tip of the tonometer does not affect the IOP values.
Purpose:
We investigated whether manual UEE affects the IOP readings using three rebound tonometers (Icare TA01i, Icare PRO, and Icare ic100) and Goldmann applanation tonometry (GAT).
Methods:
One eye was measured for 101 patients (56 eyes of primary open-angle glaucoma patients and 45 healthy subjects). The IOPs were measured without and with manual UEE. Each IOP was measured twice; the measurement order using the tonometers was randomly selected. In addition, palpebral fissure height (distance between the upper and lower eyelids) was measured.
Results:
The IOPs without manual UEE were 12.1 ± 2.9, 13.3 ± 2.7, 11.7 ± 2.9, and 16.0 ± 3.2 mmHg (Icare TA01i, Icare PRO, Icare ic100, and GAT), and those with manual UEE were 12.3 ± 3.0, 13.3 ± 2.8, 11.7 ± 2.9, and 16.0 ± 3.3, respectively. No significant difference was found between the IOP without and with manual UEE (IOP difference; all, P > .50; paired t test). Multiple linear regression analyses revealed that palpebral fissure height did not affect IOP difference for any of the tonometers.
Conclusions:
Simple manual UEE when measuring the IOP has little effect on the IOP obtained using all current rebound tonometers and GAT.
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Muscles of the Eye
Extraocular Muscles
The six extraocular muscles surround the eyeball and control its movements. They are responsible for a wide range of eye motions, including looking up, down, left, right, and...
Glaucoma: Overview