Transversal dose profile reconstruction for clinical proton beams: A detectors inter-comparison
R Catalano1, G Petringa1, G Cuttone1
1INFN-LNS, Istituto Nazionale di Fisica Nucleare - Laboratori Nazionali del Sud, Via S. Sofia, 62, 95123 Catania, Italy.
Purpose:
The main purpose of this work is the inter-comparison between different devices devoted to the transversal dose profile recostruction for daily QA tests in proton therapy.
Methods:
The results obtained with the EBT3 radiochromic films, used as a reference, and other common quality control devices, have been compared with those obtained with a beam profiling system developed at the "Laboratori Nazionali del Sud" of Italian Institute for Nuclear Physics (INFN-LNS, Catania, Italy). It consists of a plastic scintillator screen (thickness 1 mm), mounted perpendicularly to the beam axis and coupled with a highly sensitive CCD detector in a light-tight box.
Results And Conclusion:
The tests, carried out both at the INFN-LNS and Trento Proton Therapy Center facilities, show, in general, a good agreement between the different detectors. The beam profiling system, in particular, appears to be a promising quality control device for 2-D relative dosimetry, because of its linear response in a dose rate range useful for proton therapy treatments, its high spatial resolution and its short acquisition and processing time.