Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Atrial fibrillation detection with a portable device during cardiovascular screening in primary care
Adriana C Diamantino1, Bruno R Nascimento2,3, Andrea Z Beaton4
1Serviço de Cardiologia e Cirurgia Cardiovascular e Centro de Telessaúde do Serviço de Cardiologia e Cirurgia Cardiovascular e Centro de Telessaúde do, Hospital das Clinicas da Universidade Federal de Minas Gerais, Belo Horizonte, Minas Gerais, Brazil.
Introduction:
A novel handheld dual-electrode stick is a portable atrial fibrillation (AF) screening device (AFSD). We evaluated AFSD performance in primary care patients referred for echocardiogram (echo).
Methods:
The AFSD has a light indication of irregular rhythm and single-lead ECG recording. Patients were instructed to hold the device for 1 min, and AF indication was recorded. A 12-lead ECG was performed for all AFSD-positive patients and 250 patients with negative AFSD screen. Echos were performed based on a clinical risk score: all high-risk patients and a sampling of low-risk patients underwent complete echo. Intermediate risk patients first had a screening echocardiogram, with a follow-up complete study if abnormality was suspected.
Results:
In 5 days, 1518 patients underwent clinical evaluation and cardiovascular risk stratification: mean age 58±16 years, 66% women. The AFSD was positive in 6.4%: 12.6% high risk, 6.1% intermediate risk and 2.2% low risk. Older age was a risk factor (9.3% vs 4.8% in those more than and less than 65 years, p=0.001). AFSD positive was independently associated with heart disease in echo (OR=3.9, 95% CI 2.1 to 7.2, p<0.001). Compared with 12-lead ECG, the AFSD had sensitivity of 90.2% (95% CI 77.0% to 97.3%) and specificity of 84.0% (95% CI 79.3% to 88.0%) for AF detection.
Conclusion:
AFSD demonstrated high sensitivity for AF detection in primary care patients referred for echo. AF prevalence was substantial and independently associated with structural or functional heart disease, suggesting that AFSD screening could be a useful primary care tool to stratify risk and prioritise echo.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias V: Evaluating Dysrhythmias
Holter Monitor: 24-Hour Monitoring
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: