Related Experiment Video
Updated: Dec 29, 2025

Evaluating Regional Pulmonary Deposition using Patient-Specific 3D Printed Lung Models
Published on: November 11, 2020
In Vitro Study of the Effect of Breathing Pattern on Aerosol Delivery During High-Flow Nasal Therapy
Gavin Bennett1, Mary Joyce2, Louise Sweeney2
1Aerogen Limited, IDA Business Park, Dangan, Galway, Ireland. gbennett@aerogen.com.
Introduction:
The use of concurrent aerosol delivery during high-flow nasal therapy (HFNT) may be exploited to facilitate delivery of a variety of prescribed medications for inhalation. The study assessed the effect of tidal volume, breath rate, and inspiratory:expiratory (I:E) ratio on the quantity of aerosol captured at the level of the trachea during simulated HFNT.
Methods:
Testing was completed according to a factorial statistical design of experiments (DOE) approach. Tracheal dose was characterized with a vibrating mesh nebulizer (Aerogen Solo, Aerogen Ltd) using simulated adult, small child, and infant HFNT models. Furthermore, aerosol delivery was evaluated across a range of adult patient profiles with clinically representative test setups.
Results:
Aerosol delivery increased with a large tidal volume, a rapid breath rate, and a long inspiratory time. Tidal volume, breath rate, and I:E ratio each had a significant effect on tracheal dose across simulated adult, small child, and infant breathing.
Conclusion:
The main trends that were identified in the statistical DOE predicted aerosol delivery across adult patient breathing profiles, in terms of tidal volume, breath rate, and I:E ratio. Therefore, patients with distressed breathing profiles may be expected to receive a larger aerosol dose than those with normal breathing rates.
Funding:
Aerogen Limited.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...