A technique to generate synthetic CT from MRI for abdominal radiotherapy
Shu-Hui Hsu1,2, Pamela DuPre1, Qi Peng3,4
1Department of Radiation Oncology, Montefiore Medical Center, Bronx, NY, USA.
Purpose:
To investigate a method to classify tissues types for synthetic CT generation using MRI for treatment planning in abdominal radiotherapy.
Methods:
An institutional review board approved volunteer study was performed on a 3T MRI scanner. In-phase, fat and water images were acquired for five volunteers with breath-hold using an mDixon pulse sequence. A method to classify different tissue types for synthetic CT generation in the abdomen was developed. Three tissue clusters (fat, high-density tissue, and spine/air/lungs) were generated using a fuzzy-c means clustering algorithm. The third cluster was further segmented into three sub-clusters that represented spine, air, and lungs. Therefore, five segments were automatically generated. To evaluate segmentation accuracy using the method, the five segments were manually contoured on MRI images as the ground truth, and the volume ratio, Dice coefficient, and Hausdorff distance metric were calculated. The dosimetric effect of segmentation accuracy was evaluated on simulated targets close to air, lungs, and spine using a two-arc volumetric modulated arc therapy (VMAT) technique.
Results:
The volume ratio of auto-segmentation to manual segmentation was 0.88-2.1 for the air segment and 0.72-1.13 for the remaining segments. The range of the Dice coefficient was 0.24-0.83, 0.84-0.93, 0.94-0.98, 0.93-0.96, and 0.76-0.79 for air, fat, lungs, high-density tissue, and spine, respectively. The range of the mean Hausdorff distance was 3-29.1 mm, 0.5-1.3 mm, 0.4-1 mm, 0.7-1.6 mm, and 1.2-1.4 mm for air, fat, lungs, high-density tissue, and spine, respectively. Despite worse segmentation accuracy in air and spine, the dosimetric effect was 0.2% ± 0.2%, with a maximum difference of 0.8% for all target locations.
Conclusion:
A method to generate synthetic CT in the abdomen was developed, and segmentation accuracy and its dosimetric effect were evaluated. Our results demonstrate the potential of using MRI alone for treatment planning in the abdomen.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies IV: Magnetic Resonance Imaging