Assessing and Reducing Positron Emission Tomography/Computed Tomography Radiotracer Infiltrations: Lessons in Quality
Jackson W Kiser1, Thad Benefield2, Ronald K Lattanze3
1Carilion Clinic, Roanoke, VA.
Purpose:
Accurate administration of radiotracer dose is essential to positron emission tomography (PET) image quality and quantification. Misadministration (infiltration) of the dose can affect PET/computed tomography results and lead to unnecessary or inappropriate treatments and procedures. Quality control efforts ensure accuracy of the administered dose; however, they fail to ensure complete delivery of the dose into the patient's circulation. We used new technology to assess and improve infiltration rates and evaluate sustainability.
Methods:
Injection quality was measured, improved, and sustained during our participation in a multicenter quality improvement project using Define, Measure, Analyze, Improve, Control methodology. Five technologists monitored injection quality in the Measure and Improve phases. After seven new technologists joined the team in the Control phase, infiltration rates were recalculated, controlling for technologist- and patient-level correlations, and comparisons were made between these two groups of technologists.
Results:
In the Measure phase, five technologists monitored 263 injections (13.3% infiltration rate). Nonantecubital fossa injections had a higher probability of infiltration than antecubital fossa injections. After implementing a quality improvement plan (QIP), the same technologists monitored 278 injections in the Improve phase (2.9% infiltration rate). The 78% decrease in infiltration rate was significant (P < .001) as was the decrease in nonantecubital fossa infiltrations (P = .0025). In the Control phase, 12 technologists monitored 1,240 injections (3.1% infiltration rate). The seven new technologists had significantly higher rates of infiltration (P = .017).
Conclusion:
A QIP can significantly improve and sustain injection quality; however, ongoing monitoring is needed as new technologists join the team.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation I: X-ray and CT