The clinical evaluation of the new indirect calorimeter developed by the ICALIC project
Taku Oshima1, Marta Delsoglio2, Yves M Dupertuis2
1Department of Emergency and Critical Care Medicine, Chiba University Graduate School of Medicine, 1-8-1 Inohana Chuou-ku, Chiba City, Chiba, 260-8677, Japan.
Background & Aims:
The ICALIC project was initiated for developing an accurate, reliable and user friendly indirect calorimeter (IC) and aimed at evaluating its ease of use and the feasibility of the EE measurements in intensive care unit (ICU).
Methods:
This was a prospective unblinded, observational, multi-center study. Simultaneous IC measurements in mechanically ventilated ICU patients were performed using the new IC (Q-NRG®) and currently used devices. Time required to obtain EE was recorded to evaluate the ease of use of Q-NRG® versus currently used ICs and EE measurements were compared. Conventional descriptive statistics were used: data as mean ± SD.
Results:
Six centers out of nine completed the required number of patients for the primary analysis. Mean differences in the time needed by Q-NRG® against currently used ICs were -32.3 ± 2.5 min in Geneva (vs. Deltatrac®; p < 0.01), -32.3 ± 3.1 in Lausanne (vs. Quark RMR®; p < 0.05), -33.7 ± 1.4 in Brussels (vs. V-Max Encore®; p < 0.05), -26.4 ± 7.8 in Tel Aviv (vs. Deltatrac®; p < 0.05), -28.5 ± 3.5 in Vienna (vs. Deltatrac®; p < 0.05), and 0.3 ± 1.2 in Chiba (vs. E-COVX®; p = 0.17). EE (kcal/day) measurements by the Q-NRG® were similar to the Deltatrac® in Geneva and Vienna (mean differences±SD: -63.1 ± 157.8 (p = 0.462) and -22.9 ± 328.2 (=0.650)), but significantly different in Tel Aviv (307.4 ± 324.5, p < 0.001). Significant differences were observed in Lausanne (Quark RMR®: -224.4 ± 514.9, p = 0.038) and in Brussels (V-max®: -449.6 ± 667.4, p < 0.001), but none was found in Chiba (E-COVX®; 55.0 ± 204.1, p = 0.165).
Conclusion:
The Q-NRG® required a much shorter time than most other ICs to determine EE in mechanically ventilated ICU patients. The Q-NRG® is the only commercially available IC tested against mass spectrometry to ensure gas accuracy, while being very easy-to use.
Related Concept Videos
Calorimetry
Constant Pressure Calorimetry
Constant Volume Calorimetry
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...