An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
Assessment of the Antithrombogenicity of a Poly-2-Methoxyethylacrylate-Coated Central Venous Port-Catheter System
Shuji Kariya1, Miyuki Nakatani2, Yasuyuki Ono2
1Department of Radiology, Kansai Medical University, 2-5-1 Shinmachi, Hirakata, Osaka, 5731010, Japan. kariyas@hirakata.kmu.ac.jp.
Purpose:
The purpose of this study was to evaluate the antithrombogenic effects of poly-2-methoxyethylacrylate (PMEA) coating applied to the internal surfaces of the port-catheter system to prevent thrombotic occlusion.
Materials And Methods:
PMEA coating was applied to the inner surfaces of the entire system, including the chamber, catheter, and connecting stem. PMEA-coated and uncoated port-catheter systems were each filled with fresh human blood for 1 or 3 h and then flushed with saline. Volumes of residual thrombi in the system and protein in the catheter were then compared.
Results:
Saline flushing of the PMEA-coated port-catheter system expelled all visually recognizable thrombi, leaving no unremovable adhesions. In contrast, the uncoated port showed thrombi adherent to the inner surfaces and incomplete expulsion with flushing. Mean (± standard deviation) residual thrombus mass was significantly lower in PMEA-coated port-catheter systems (20.5 ± 6.2 mg) than in uncoated systems (230.3 ± 92.5 mg; p < 0.01). Mean residual protein was likewise significantly lower in PMEA-coated systems (20.5 ± 6.2 mg) than in uncoated systems (230.3 ± 92.5 mg; p < 0.01).
Conclusions:
PMEA coating on a central venous port-catheter system decreased accumulation and facilitated flushing of thrombi from the system by flushing, as compared with the uncoated system. PMEA-coated central venous port-catheter systems appear relatively antithrombotic compared to uncoated systems.
Insights
Poly-2-methoxyethylacrylate (PMEA) coating significantly reduced thrombus formation in port-catheter systems. This antithrombotic coating facilitates easier flushing of residual thrombi and protein, improving system performance.
Area of Science:
- Biomaterials Science
- Medical Device Engineering
- Thrombosis Research
Background:
- Central venous port-catheter systems are prone to thrombotic occlusion, a complication that can impair device function and lead to adverse events.
- Preventing thrombus formation on internal surfaces is crucial for maintaining the patency and longevity of these medical devices.
Purpose of the Study:
- To investigate the antithrombogenic efficacy of poly-2-methoxyethylacrylate (PMEA) coating on the internal surfaces of port-catheter systems.
- To assess the PMEA coating's ability to prevent thrombotic occlusion and facilitate thrombus removal.
Main Methods:
- PMEA coating was applied to the internal surfaces of port-catheter systems (chamber, catheter, connecting stem).
- Coated and uncoated systems were filled with fresh human blood for 1 or 3 hours.
- Residual thrombus mass and protein levels were quantified after saline flushing and compared between coated and uncoated systems.
Main Results:
- PMEA-coated systems showed complete expulsion of visually recognizable thrombi with saline flushing, leaving no adhesions.
- Uncoated systems exhibited adherent thrombi and incomplete expulsion upon flushing.
- Significantly lower mean residual thrombus mass (20.5 ± 6.2 mg vs. 230.3 ± 92.5 mg; p < 0.01) and protein levels were observed in PMEA-coated systems compared to uncoated systems.
Conclusions:
- PMEA coating effectively reduces thrombus accumulation within central venous port-catheter systems.
- The antithrombotic properties of PMEA facilitate easier flushing of thrombi and proteins, enhancing system performance.
- PMEA-coated port-catheter systems demonstrate a significant antithrombotic advantage over uncoated systems.