Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Inter- and Intraobserver Agreement in Measuring Urolithiasis Density on Nonenhanced Computed Tomography
Guus W Disselhorst1, Jorik J Reimerink2, Rogier K van der Vijgh3
1Department of Urology, OLVG Hospital, Amsterdam, The Netherlands.
Standardizing Hounsfield unit (HU) measurements for kidney stones using a specific protocol significantly improves observer agreement, especially for urologists. This enhances treatment planning for urolithiasis.
Area of Science:
- Radiology
- Urology
- Medical Imaging
Background:
- Treatment decisions for urolithiasis rely partly on kidney stone density measured in Hounsfield units (HU) via nonenhanced computed tomography (NECT).
- Variability among observers in HU measurements can impact treatment choices.
- This study investigates observer agreement in HU measurements and the effect of a standardized protocol.
Purpose of the Study:
- To assess inter- and intraobserver agreement for HU measurements in NECT of urolithiasis.
- To determine if a specific measurement protocol improves agreement.
- To identify the optimal protocol for consistent HU density assessment.
Main Methods:
- Retrospective analysis of 155 NECT scans from patients with kidney stones (≥4 mm).
- Five observers (radiologists, urologists, and residents) performed measurements four times each.
- Measurements were taken without a protocol and with two different protocols (A and B) involving bone setting, zoom, and specific plane selection.
Main Results:
- Initial interobserver agreement without a protocol was good (ICC=0.84).
- Protocol A decreased agreement (ICC=0.62), while Protocol B significantly improved it (ICC=0.90).
- Protocol B also enhanced intraobserver agreement (ICC=0.93) and showed the largest improvement for urologists.
Conclusions:
- Good observer agreement for HU measurements in urolithiasis exists without a protocol.
- A protocol using zoom, bone setting, and measurement in a representative plane (Protocol B) is recommended for improved consistency.
- Measuring in a single representative plane is effective, and measuring in three planes does not further increase agreement.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies II: Ultrasonography
Imaging Studies VI: Voiding Cystourethrography and Cystography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT