The effect of flow and pressure on the intraoxygenator flow path of different contemporary oxygenators: an in vitro
Rik Hj Hendrix1, Alexander Ks Yeung1, Yuri M Ganushchak1
1Department of Extra-Corporeal Circulation, Maastricht University Medical Centre, Maastricht, The Netherlands.
Introduction:
This study analyzed the effect of different flows and pressures on the intraoxygenator flow path in three contemporary oxygenators and its consequences for oxygen transfer efficiency.
Methods:
In an experimental setup, intraoxygenator flow path parameters were analyzed at post-oxygenator pressures of 150, 200, and 250 mm Hg and at flows ranging from 2 L/min to the oxygenators' maximum permitted flow, with and without pulsatility. The oxygen gradient and the oxygen transfer per minute and per 100 mL blood were calculated using previously collected clinical data and compared with the flow path parameters.
Results:
Increasing pressure did not affect the flow path parameters, whereas pulsatile flow led to significantly increased dynamic oxygenator blood volumes. Increased flow resulted in decreased values of the flow path parameters in all oxygenators, indicating increased flow through short pathways in the oxygenator. In parallel, oxygen transfer/100 mL blood decreased in all oxygenators (average 2.5 ± 0.4 to 2.4 ± 0.3 mL/dL, p > 0.001) and the oxygen gradient increased from 229 ± 45 to 287 ± 29 mm Hg, p > 0.001, indicating decreased oxygen transfer efficiency. Oxygen transfer/min increased (101 ± 15 to 143 ± 20 mL/min/m2, p > 0.001), however, due to the increased flow through the oxygenator.
Conclusion:
Varying trans-membrane oxygenator pressures did not lead to changes in the intraoxygenator flow path, while an increased flow exhibited lower flow path parameters resulting in less efficient use of the gas exchange compartment. The latter was confirmed by a decrease in O2 transfer efficiency during higher blood flows.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Transport in the Blood
General External Flow Characteristics
Gradually Varying Flow
Design Example: Flow of Oil Through Circular Pipes