Optometry students showed lower repeatability in masked Goldmann tonometry compared to experienced optometrists, especially with higher intraocular pressure. Further training with complex cases is recommended for improved glaucoma patient eye pressure measurement.
Area of Science:
- Ophthalmology
- Optometry
- Glaucoma Research
Background:
- Goldmann tonometry is a standard method for measuring intraocular pressure (IOP).
- Repeatability of tonometry measurements is crucial for accurate glaucoma diagnosis and management.
- Assessing the performance of optometry students in clinical procedures is essential for training quality.
Purpose of the Study:
- To evaluate the repeatability of masked Goldmann tonometry.
- To compare the performance of optometry students (interns) with experienced optometrists.
- To identify factors influencing tonometry measurement variability in glaucoma patients.
Main Methods:
- Masked Goldmann tonometry was performed by two fourth-year optometry interns and one attending optometrist on glaucoma patients.
- Each examiner conducted three sequential measurements on each eye.
- Data analysis included coefficient of variation and interobserver differences.
Main Results:
- Interns exhibited lower intraobserver and interobserver repeatability (CV: 9.1-12.1%) compared to optometrists (CV: 6.4-6.6%).
- Intern measurements showed increased variability and overestimation at higher IOP levels, influenced by both intern and patient factors.
- Optometrist measurements were more consistent and not affected by IOP levels.
Conclusions:
- Optometry student performance in masked Goldmann tonometry is less repeatable than that of experienced optometrists.
- Variability in intern measurements increases with higher intraocular pressure and is influenced by patient-specific factors.
- Enhanced clinical exposure to patients with elevated IOP and further research on factors affecting trainee tonometry are necessary.