An assessment of the quality of optical coherence tomography image acquisition
Elder Iarossi Zago1, Abdul Jawwad Samdani2, Gabriel Tensol Rodrigues Pereira2
1Cardiovascular Imaging Core Laboratory, Harrington Heart & Vascular Institute, University Hospitals Cleveland Medical Center, 11100 Euclid Avenue, Lakeside Building, Room 3113, Mailstop: Lakeside 5038, Cleveland, OH, 44106, USA. elderzago@gmail.com.
Abstract:
Optical coherence tomography (OCT) provides excellent image resolution, however OCT optimal acquisition is essential but could be challenging owing to several factors. We sought to assess the quality of OCT pullbacks and identify the causes of suboptimal image acquisition. We evaluated 784 (404 pre-PCI; 380 post-PCI) coronary pullbacks from an anonymized OCT database from our Cardiovascular Imaging Core Laboratory. Imaging of the region-of-interest (ROI-lesion or stented segment plus references) was incomplete in 16.1% pullbacks, caused by pullback starting too proximal (63.7%), inappropriate pullback length (17.1%) and pullback starting too distal (11.4%). The quality of image acquisition was excellent in 36.3% pullbacks; whereas 4% pullbacks were unanalyzable. Pullback quality was most commonly affected by poor blood displacement from inadequate contrast volume (27.4%) or flow (25.6%), followed by artifacts (24.1%). Acquisition mode was 'High-Resolution' (54 mm) in 74.4% and 'Survey' (75 mm) in 25.6% of cases. The 54 mm mode was associated with incomplete ROI imaging (p = 0.020) and inadequate contrast volume (p = 0.035). We observed a substantial frequency of suboptimal image acquisition and identified its causes, most of which can be addressed with minor modifications during the procedure, ultimately improving patient outcomes.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...