Related Experiment Video
Updated: Dec 28, 2025

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
[Measurement Accuracy of CT Beam Width with a Portable Flat-panel Detector]
Nao Ichikawa1,2, Kosuke Matsubara3, Atsushi Fukuda4
1Department of Radiation Oncology, Shiga General Hospital.
Portable flat-panel detector (FPD) systems accurately measure computed radiography (CR) CT beam width. This study confirms FPDs offer equivalent measurement accuracy to traditional CR systems, supporting their clinical adoption for CT beam width assessment.
Area of Science:
- Medical Imaging Physics
- Radiological Technology
Background:
- Computed radiography (CR) systems have been standard for clinical X-ray imaging.
- Portable flat-panel detector (FPD) systems are increasingly replacing traditional CR devices.
- FPD systems present a potential alternative for measuring CT beam width.
Purpose of the Study:
- To evaluate the accuracy of portable flat-panel detector (FPD) systems for measuring computed tomography (CT) beam width.
- To compare the measurement accuracy of FPD systems against traditional CR systems.
Main Methods:
- A CR cassette and an FPD were positioned at the isocenter for axial CT scans.
- Pixel values were measured under varying tube currents (10-40 mA) and a fixed tube potential (80 kVp).
- A double exposure technique with copper and lead shielding was used to measure beam widths (40-160 mm) and prevent saturation/artifacts.
Main Results:
- Log-linear relationships were established between pixel values and radiation exposure (up to 12.5 mAs) for both CR and FPD systems.
- Statistical analysis demonstrated equivalent measurement accuracy between the CR and FPD systems.
- Confidence intervals confirmed the non-inferiority of FPD measurements.
Conclusions:
- Portable FPD systems are suitable for measuring CT beam width.
- The accuracy of FPD systems is comparable to that of CR systems for this application.
- FPD systems can be reliably used as a substitute for CR in CT beam width measurements.
More Related Videos
10:39Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating
Published on: October 11, 2016
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016