Related Experiment Video
Updated: Dec 28, 2025

Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
Published on: June 20, 2015
First-in-human continuous 24-hour measurement of intraocular pressure and ocular pulsation using a novel contact lens
Robert Wasilewicz1, Thierry Varidel2, Sonja Simon-Zoula2
1[WASILEWICZ]-Eye Clinic, Poznan, Poland.
Background/Aims:
This study assessed the feasibility of a novel contact lens device for intraocular pressure (IOP) and ocular pulse amplitude (OPA) continuous measurements over 24 hours.
Methods:
This prospective, open-label, single-centre, non-randomised study included glaucoma and healthy subjects. IOP and OPA values acquired by the pressure-measuring contact lens (PMCL) device in one patient's eye at the beginning of the measurement were compared with tonometry values (Goldman applanation tonometry (GAT) and dynamic contour tonometry (DCT)) in the same eye just before PMCL placement. Furthermore, IOP and OPA values measured with PMCL on the study eye during a water drinking test (WDT) were compared with DCT values in the fellow eye. Comparisons were performed using t-tests with 95% Confidence Intervals.
Results:
Twenty-four-hour IOP and OPA curves were obtained for eight subjects. The mean IOP difference between PMCL and tonometry on the same eye was within ±5 mm Hg in 75% (GAT) and 87.5% (DCT) of subjects. IOP variations due to WDT were detected by PMCL and DCT, showing an average increase of 2.43 and 1.85 mm Hg, respectively. Differences between PMCL and DCT for IOP variations in fellow eyes were within ±5 mm Hg for 97.2% of time points. The difference between OPA in fellow eyes was within ±5 mm Hg for 85.5% of the time points.
Conclusions:
This first-in-human study is a proof-of-concept for 24-hour continuous measurements of IOP and OPA with the PMCL. This device is non-invasive and has good comparability with standard tonometry.