Changes in peak inspiratory flow rate and peak airway pressure with endotracheal tube size during chest compression
Jung Wan Kim1, Jin Woong Lee1, Seung Ryu1
1Department of Emergency Medicine, Chungnam National University Hospital, Jung-gu, Daejeon, Republic of Korea.
Background:
Adequate airway management plays an important role in high-quality cardiopulmonary resuscitation (CPR). Airway management is usually performed using an endotracheal tube (ETT) during CPR. However, no study has assessed the effect of ETT size on the flow rate and airway pressure during CPR.
Methods:
We measured changes in peak inspiratory flow rate (PIFR), peak airway pressure (Ppeak), and mean airway pressure (Pmean) according to changes in ETT size (internal diameter 6.0, 7.0, and 8.0 mm) and with or without CPR. A tidal volume of 500 mL was supplied at a rate of 10 times per minute using a mechanical ventilator. Chest compressions were maintained at a constant compression depth and speed using a mechanical chest compression device (LUCAS2, mode: active continuous, chest compression rate: 102±2/minute, chest compression depth 2-2.5 inches).
Results:
The median of several respiratory physiological parameters during CPR was significantly different according to the diameter of each ETT (6.0 vs. 8.0 mm): PIFR (32.1 L/min [30.5-35.3] vs. 28.9 L/min [27.5-30.8], P<0.001), Ppeak (48.84 cmH2O [27.46-52.11] vs. 27.45 cmH2O [22.53-52.57], P<0.001), and Pmean (18.34 cmH2O [14.61-21.66] vs.13.66 cmH2O [8.41-19.24], P<0.001).
Conclusion:
The changes in PIFR, Ppeak, and Pmean were related to the internal diameter of ETT, and these values tended to decrease with an increase in ETT size. Higher airway pressures were measured in the CPR group than in the no CPR group.
Related Concept Videos
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...