Improvement of performance of minimum variance beamformer by introducing cross covariance estimate
Hideyuki Hasegawa1,2, Ryo Nagaoka3
1Faculty of Engineering, Academic Assembly, University of Toyama, 3190 Gofuku, Toyama, 930-8555, Japan. hasegawa@eng.u-toyama.ac.jp.
Purpose:
The delay-and-sum beamformer is widely used in clinical ultrasound systems to obtain ultrasonic images. To improve image quality, the minimum variance (MV) beamformer was introduced in medical ultrasound imaging. The MV beamformer determines beamformer weights from ultrasonic echo signals received by individual transducer elements in an ultrasonic probe. In the present study, the MV beamformer was investigated to improve its performance.
Methods:
In MV beamforming, a covariance matrix of echo signals received by individual elements needs to be estimated to obtain adaptive beamformer weights. To obtain a stable estimate, a total receiving aperture is divided into subarrays, and a covariance matrix is obtained using echo signals from each subarray to average covariance matrices from all subarrays. This procedure is called "subarray averaging." In the present study, a new method for estimation of the covariance matrix was proposed. In the proposed method, a covariance matrix, namely, a cross covariance matrix, is obtained using echo signals from different subarrays. Multiple covariance matrices are obtained from all different pairs of subarrays and averaged.
Results:
In the present study, the performance of the proposed method was evaluated by basic experiments on a phantom. Lateral spatial resolutions obtained by MV beamforming with conventional subarray averaging and the proposed method were similar. However, contrast obtained by MV beamforming with the proposed method was - 0.56 dB, which was significantly better than the - 5.06 dB obtained by MV beamforming with conventional subarray averaging.
Conclusion:
Image contrast in MV beamforming could be improved significantly by estimating "cross" covariance matrices.
More Related Videos
10:39Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating
Published on: October 11, 2016
09:01Gain-compensation Methodology for a Sinusoidal Scan of a Galvanometer Mirror in Proportional-Integral-Differential Control Using Pre-emphasis Techniques
Published on: April 4, 2017
Related Concept Videos
Linear Approximation in Frequency Domain
In contrast, nonlinear systems do not inherently possess these properties. However, for small deviations around an operating point, a nonlinear system can often be approximated as linear....
Residuals and Least-Squares Property
If the observed data point lies above the line, the residual is positive, and the line underestimates the actual data value for y. If the observed data point lies below the line, the residual is negative, and the line overestimates the actual data value for y.
The process of fitting the best-fit...
Variance
The standard deviation measures the spread in the same units as the data....
Linear Approximation in Time Domain
For a simple pendulum with a mass evenly distributed along its length and the center of mass located at half the pendulum's length,...
Calibration Curves: Correlation Coefficient
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...