Real-time estimation of mean arterial blood pressure based on photoplethysmography dicrotic notch and perfusion
Jona Joachim1,2,3, Maxime Coutrot4,5,6, Sandrine Millasseau7
1Department of Anesthesiology, Burn and Critical Care, St-Louis-Lariboisière-Fernand Widal University Hospitals, Assistance Publique - Hôpitaux de Paris, 2 rue Ambroise Paré, 75010, Paris, France. jona.joachim@aphp.fr.
Abstract:
Hypotension during general anesthesia is associated with poor outcome. Continuous monitoring of mean blood pressure (MAP) during anesthesia is useful and needs to be reliable and minimally invasive. Conventional cuff measurements can lead to delays due to its discontinuous nature. It has been shown that there is a relationship between MAP and photoplethysmography (PPG) parameters like the dicrotic notch and perfusion index (PI). The objective of the study was to continuously estimate MAP from PPG. Pulse wave analysis based on PPG was implemented using either notch relative amplitude (MAPNRA), notch absolute amplitude (MAPNAA) or PI (MAPPI) to estimate MAP from PPG waveform features during general anesthesia. Estimated MAP values were compared to brachial cuff MAP (MAPcuff) and to radial invasive MAP (MAPinv). Forty-six patients were analyzed for a total of 235 h. Compared to MAPcuff, mean bias and limits of agreement were 1 mmHg (- 26 to +29), - 1 mmHg (- 10 to +8) and - 3 mmHg (- 21 to +13) for MAPNRA, MAPNAA and MAPPI respectively. Compared to MAPinv, mean absolute error (MAE) was 20 mmHg [10 to 39], 11 mmHg [5 to 18] and 16 mmHg [9 to 24] for MAP derived from MAPNRA, MAPNAA and MAPPI respectively. When calibrated every 5 min, MAPNAA showed a MAE of 6 mmHg [5 to 9]. MAPNAA provides the best estimates with respect to brachial cuff MAP and invasive MAP. Regular calibration allows to reduce drift over time. Beat to beat estimation of MAP during general anesthesia from the PPG appears possible with an acceptable average error.
More Related Videos
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.