Related Experiment Video
Updated: Dec 28, 2025

Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification
Published on: September 21, 2011
Comparison of polysaccharide-based and protein-based chiral liquid chromatography columns for enantioseparation of
Yandi Fu1, Francesc Borrull1, Núria Fontanals1
1Department of Analytical Chemistry and Organic Chemistry, Universitat Rovira i Virgili, Tarragona, Spain.
Abstract:
Two different columns-Lux Cellulose-1 and Chiralpak CBH-were evaluated for their chiral recognition abilities for eight drugs comprising three β-blockers, one antacid, and four cathinones in polar-organic elution mode and reversed-phase elution mode, respectively. The factors that affected the enantioseparation were tested and optimized to develop a suitable chiral separation method whose LC conditions are compatible with MS detection. In polar-organic elution mode with the Lux Cellulose-1 column, methanol and acetonitrile were tested as the main components of the mobile phase. In addition, the effects of adding isopropanol as organic modifier, acidic additives (formic acid), and basic additives (diethylamine) were evaluated. In reversed-phase elution mode with the Chiralpak CBH column, the effect of type and concentration of organic modifier (isopropanol, acetonitrile, and methanol), the mobile phase pH (6.4 and 5.0), and buffer concentration (1mM-20mM ammonium acetate) were evaluated. The best enantioseparation was achieved with the Chiralpak CBH column with a mobile phase composed of 5mM ammonium acetate aqueous (pH = 6.4)/methanol (95/5, v/v) at a flow rate of 0.1 mL/min and a temperature of 30°C. Under these conditions, six of eight chiral drugs were baseline separated.
Related Concept Videos
Ion-Exchange Chromatography
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Principles Of Column Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...