Related Experiment Video
Updated: Dec 28, 2025

Sound Source Localization Testing in Single-sided Deafness Following Bone Conduction Intervention
Published on: December 20, 2024
Active Transcutaneous Bone Conduction Implant: Audiometric Outcomes Following a Novel Middle Fossa Approach With
Lauren Siegel1, Peng You1, Kim Zimmerman2
1Department of Otolaryngology-Head and Neck Surgery, Schulich School of Medicine and Dentistry, Western University.
Objective:
To present surgical and audiometric outcomes of patients implanted with an active transcutaneous bone conduction implant following the novel middle fossa surgical approach with self-drilling screws.
Study Design:
Retrospective review.
Setting:
Tertiary care center.
Patients:
Thirty-seven adults with either conductive or mixed hearing loss that met indications for an active transcutaneous bone conduction implant were consecutively implanted from April, 2013 to May, 2018.
Intervention:
Unilateral middle fossa implantation of an active transcutaneous bone conduction implant.
Main Outcome Measures:
Patient charts were reviewed for surgical outcomes and complications over the 6-year period. Preoperative air conduction, preoperative bone conduction, and 3-month postoperative aided thresholds were recorded. Speech perception was assessed using CNC words and AzBio sentences. Pure-tone averages (PTAs; measured at 0.5, 1.0, 2.0 and 3.0 kHz), air-bone gap, and functional gain were calculated.
Results:
Mean air conduction and bone conduction PTAs (±standard deviation) of the implanted ear were 66.8 dB (±14.9 dB) and 21.9 dB (±14.0 dB), respectively. Mean aided PTA was 26.5 dB (± 8.5 dB). The average functional gain was 40.3 dB (±19.0 dB). Favorable speech perception outcomes were observed. No complications or instances of revision surgery were reported, with a mean follow-up time of 32 months (range, 9-71 mo).
Conclusions:
This is the first paper to describe outcomes of patients implanted with an active transcutaneous bone conduction implant via the middle fossa with self-drilling screws. Favorable surgical outcomes were observed with a follow-up of up to 6 years.