Background:
Vitamin A and E are routinely monitored to assess nutritional status. The most commonly used approach for their measurement involves laborious liquid-liquid extraction followed by high-performance liquid chromatography (HPLC) analysis on dedicated instrumentation. We describe a simple, rapid protocol for measurement of vitamin A and E and their integration into an existing online sample preparation liquid chromatography tandem mass spectrometry (SPLC-MS/MS) workflow.
Methods:
We performed a method comparison between the SPLC-MS/MS and HPLC methods for vitamin A and E by measuring patient specimens across the concentration range 11-81 µg/dL for vitamin A and 1-18 mg/L for vitamin E. The analysis times on each platform were also compared.
Results:
SPLC-MS/MS and HPLC methods were comparable with regards to analytical performance; mean bias across the measured range was 2.54% (95% CL: -11.56-16.64%) for vitamin A and -2.04% (95% CL: -18.20-14.12%) for vitamin E. Total analysis times were 7 min and 15 min for SPLC-MS/MS and HPLC respectively.
Conclusions:
The development of a simplified sample preparation protocol and the use of multiplexing SPLC-MS/MS have reduced sample analysis times for vitamin A and E. This method has also optimized clinical workflow through consolidation of previously independent benches.
More Related Videos
05:03Author Spotlight: Investigating Physiological Functions of Vitamin A Transporters Using HPLC-Based Vitamin A Profiling
Published on: December 27, 2024
11:09An HS-MRM Assay for the Quantification of Host-cell Proteins in Protein Biopharmaceuticals by Liquid Chromatography Ion Mobility QTOF Mass Spectrometry
Published on: April 17, 2018
Related Concept Videos
Tandem Mass Spectrometry
MALDI-TOF Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...