New microvascular anastomotic device for end-to-side anastomosis using negative pressure; a preliminary study
Kazufumi Tachi1, Katsuko S Furukawa2,3,4, Isao Koshima1
1Plastic and Reconstructive Surgery, Graduate School of Medicine, University of Tokyo, Tokyo, Japan.
Journal of Plastic Surgery and Hand Surgery
|February 25, 2020
Summary
This study explored using negative pressure devices for end-to-side vascular anastomosis in rats. While demonstrating feasibility, the low patency rate indicates a need for further device development.
Area of Science:
- Biomedical Engineering
- Surgical Innovation
- Vascular Surgery
Background:
- Negative pressure has been successfully applied to end-to-end anastomosis.
- End-to-side anastomosis is a common and critical surgical procedure.
Purpose of the Study:
- To develop and evaluate a negative pressure device for end-to-side microvascular anastomosis.
- To assess the feasibility and patency of negative pressure-assisted end-to-side anastomosis in an animal model.
Main Methods:
- A novel negative pressure device was fabricated using laser lithography and 3D printing.
- End-to-side anastomosis between the superficial epigastric vein and femoral vein was performed in nine rats.
- Patency was assessed macroscopically and microscopically at immediate post-procedure and one-week follow-up.
Main Results:
- Eight out of nine anastomoses were patent immediately after the procedure.
- Two out of nine anastomoses remained patent at the one-week follow-up.
- Vessel dislocation due to clamp loosening was identified as a cause of failure.
Conclusions:
- Negative pressure can be applied to facilitate end-to-side microvascular anastomosis.
- The current device design requires further refinement to improve patency rates.
- This technology holds potential for semi-automated vascular reconstructions.
Keywords:
Devicelymphatico-venular anastomosis (LVA)microvascular anastomosisnegative pressure vacuum