Development of an Automatic pH Adjustment Instrument for the Preparation of Analytical Samples Prior to Solid Phase
Yanbei Zhu1, Sayaka Nishigori2, Nobuaki Shimura2
1National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), 1-1-1 Umezono, Tsukuba, Ibaraki, 305-8563, Japan. yb-zhu@aist.go.jp.
Abstract:
An automatic pH adjustment instrument was developed for the preparation of analytical samples prior to solid phase extraction, which is widely used as a pretreatment technique for the separation of sample matrixes and preconcentration of elements for analysis. Real-time monitoring of the sample pH condition was performed by observing the light signal intensity of the pH-sensitive wavelength of the pH indicating reagent. A light of pH-insensitive wavelength was selected as the reference light to cancel the signal intensity variation of the pH-sensitive light due to the difference of pH indicating reagent concentration, possible difference in transparency of sample vessels, and minute fluctuation of the light source. The pH condition was elevated by automatic addition of ammonia solution using a nebulizer in the flow line of which an electromagnetic valve was equipped. Open and close operation of the electromagnetic valve was controlled based on the difference between the real-time pH condition and the target pH condition. The effectiveness of the instrument was confirmed by using various pH indicating reagents and by analyzing trace elements in a seawater certified reference material.
Related Concept Videos
Sample Preparation for Analysis: Overview
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Development of Analytical Methods
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...