Effects of Patient Size and Radiation Dose on Iodine Quantification in Dual-Source Dual-Energy CT
Xia Jiang1, Xiangyu Yang1, David E Hintenlang1
1Department of Radiology, Ohio State University College of Medicine, 395 W 12th Ave, Columbus, OH 43210.
Rationale And Objectives:
The purpose of this study was to investigate the potential effects of patient size and radiation dose on the accuracy of iodine quantification using dual-source dual-energy computed tomography (CT).
Materials And Methods:
Three phantoms representing different patient sizes were constructed, containing iodine inserts with concentrations from 0 to 20 mg/ml. Dual-energy CT scans were performed at six dose levels from 2 to 30 mGy. Iodine concentrations were measured using a three-material-decomposition algorithm and their accuracy was assessed.
Results:
In a small phantom, iodine quantification was accurate and consistent at all dose levels. In a medium phantom, minor underestimations were observed, and the results were consistent except at low dose. In the large phantom, more significant underestimation of iodine concentration was observed at higher doses (≥15 mGy), which was attributed to the beam-hardening effect. At lower doses, increasing upward bias was observed in the CT number, leading to significant overestimations of both iodine concentration and fat fraction, which was attributed to the photon-starvation effect. The severity of the latter effect was determined by mA instead of mAs, suggesting that the electronic noise, rather than the quantum noise, was responsible for the bias. Using higher kVp for the low-energy tube was found to alleviate these effects.
Conclusion:
Reliable iodine quantification can be achieved using dual-source CT, but the result can be affected by patient size and dose rate. In large patients, biases may occur due to the beam-hardening and the photon-starvation effects, in which case higher dose rate and higher kVp are recommended to minimize these effects.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET