A step and shoot intensity modulated technique for total body irradiation
Lotte S Fog1, Vibeke N Hansen2, Flemming Kjær-Kristoffersen1
1Dept. of Oncology, Rigshospitalet, University of Copenhagen, Denmark.
Technical Innovations & Patient Support in Radiation Oncology
|February 26, 2020
Summary
Step and shoot intensity-modulated radiation therapy (SS IMRT) is a feasible technique for total body irradiation (TBI) conditioning before bone marrow transplant. This method delivers targeted radiation doses, demonstrating its efficacy in clinical settings.
Area of Science:
- Radiation Oncology
- Medical Physics
Background:
- Total body irradiation (TBI) is a critical component of conditioning regimens for bone marrow transplantation.
- Conventional TBI techniques may pose challenges in dose delivery and organ-at-risk sparing.
Purpose of the Study:
- To introduce and evaluate a step and shoot intensity-modulated radiation therapy (SS IMRT) technique for TBI.
- To assess the feasibility and dosimetric accuracy of SS IMRT for TBI in clinical practice.
Main Methods:
- Patients underwent CT scanning with 5 mm slices for treatment planning.
- Multiple multi-leaf collimator (MLC) defined fields were employed in the SS IMRT technique.
- In-vivo dosimetry was conducted on 113 patients, and calculated doses for 22 patients were analyzed.
Main Results:
- In-vivo dosimetry at Royal Marsden showed a mean TLD measured dose difference of -1.9% (SD 4.5%).
- SS IMRT plans for Rigshospitalet patients achieved mean doses of 11.1-11.8 Gy to the brain, lungs, and kidneys.
- Organ-at-risk sparing was demonstrated, with V(12 Gy) below 5% for the brain, 2% for the lungs, and 0% for the kidneys.
Conclusions:
- Step and shoot IMRT is a feasible technique for delivering total body irradiation.
- SS IMRT enables precise dose delivery to target volumes while sparing organs at risk.
- This technique can be implemented using standard external beam radiation equipment.