Related Experiment Video
Updated: Dec 27, 2025

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
A photographic technique for quick assessment of mechanical isocenter of a linear accelerator
Janne Heikkilä1, Jan Seppälä1, Jouni Nuutinen1
1Cancer Center, Kuopio University Hospital, Kuopio, Finland.
Abstract:
Demands for mechanical accuracy of medical linear accelerators are increased due to the stereotactic and modulated rotational treatments. Mechanical inaccuracies affect the size and shape of the mechanical and radiation isocenters. In practice, the mechanical isocenter is defined by the intersection of rotational axes. However, there are no simple tools to check the properties of the mechanical isocenter in 3D. We introduce a new photography-based method for quick and sub-millimeter accurate determination of the mechanical isocenter. The method is based on image-processing algorithm and modified front pointer. The results demonstrate the quick measurement and visualization of the mechanical isocenter.

