Patient motion tracking for non-isocentric and non-coplanar treatments via fixed frame-of-reference 3D camera
Sergey Gasparyan1, Kyle Ko1, Lawrie B Skinner1
1Department of Radiation Oncology, Stanford university, Palo Alto, CA, 94304, USA.
Purpose:
As C-arm linac radiation therapy evolves toward faster, more efficient delivery, and more conformal dosimetry, treatments with increasingly complex couch motions are emerging. Monitoring the patient motion independently of the couch motion during non-coplanar, non-isocentric, or dynamic couch treatments is a key bottleneck to their clinical implementation. The goal of this study is to develop a prototype real-time monitoring system for unconventional beam trajectories to ensure a safe and accurate treatment delivery.
Methods:
An in-house algorithm was developed for tracking using a couch-mounted three-dimensional (3D) depth camera. The accuracy of patient motion detection on the couch was tested on a 3D printed phantom created from the body surface contour exported from the treatment planning system. The technique was evaluated against a commercial optical surface monitoring system with known phantom displacements of 3, 5, and 7 mm in lateral, longitudinal, and vertical directions by placing a head phantom on a dynamic platform on the treatment couch. The stability of the monitoring system was evaluated during dynamic couch trajectories, at speeds between 10.6 and 65 cm/min.
Results:
The proposed monitoring system agreed with the ceiling mounted optical surface monitoring system in longitudinal, lateral, and vertical directions within 0.5 mm. The uncertainty caused by couch vibration increased with couch speed but remained sub-millimeter for speeds up to 32 cm/min. For couch speeds of 10.6, 32.2, and 65 cm/min, the uncertainty ranges were 0.27- 0.73 mm, 0.15-0.87 mm, and 0.28-1.29 mm, respectively.
Conclusion:
By mounting a 3D camera in the same frame-of-reference as the patient and eliminating dead spots, this proof of concept demonstrates real-time patient monitoring during couch motion. For treatments with non-coplanar beams, multiple isocenters, or dynamic couch motion, this provides additional safety without additional radiation dose and avoids some of the complexity and limitations of room mounted systems.
Related Concept Videos
Relative Motion Analysis using Rotating Axes
However, to express the relative position of point B relative to point A, an additional frame of reference, denoted as x'y', is necessary. This additional frame not only translates but also rotates relative to the fixed frame, making it...
Non-inertial Frames of Reference
Inertial Frames of Reference
Absolute Motion Analysis- General Plane Motion
As the drone's propellers rotate, an upward force is generated that counteracts the force of gravity, enabling the drone to lift off from the ground. This initial movement of the drone is along a straight path, representing a form of translational motion. In this phase, every point on the...
Relative Motion Analysis using Rotating Axes-Problem Solving
Here, in order to determine the magnitude of velocity and acceleration for point...