The optimal use of automated office blood pressure measurement in clinical practice
Emmanuel A Andreadis1, Charalampia V Geladari1, Epameinondas T Angelopoulos1
1Hypertension and Cardiovascular Disease Prevention Center, Athens Medical Group, Psychiko Clinic, Athens, Greece.
Abstract:
This evidence-based article endorses the use of automated office blood pressure (AOBP). AOBP is the most favorable office blood pressure (BP) measuring technique as it provides accurate readings with 3-15 mm Hg lower values than the casual conventional office measurements with auscultatory or semi-automated oscillometric devices and relates closely to awake ABP readings. The AOBP technique seems to be superior to conventional office BP in predicting hypertension-mediated organ damage and appears to be equally reliable to awake ABP in the prediction of cardiovascular (CV) disease. AOBP readings should be obtained either unattended, with the patient alone in the examination room, or attended with the presence of personnel in the room but with no talking to the patient, although this recommendation is not frequently followed in routine clinical practice. To optimize office BP readings, the type of device, the rest period before AOBP measurements (preceding rest), and the time intervals between measurements were evaluated. As AOBP readings have the advantage of removing many confounding factors, the authors propose to perform measurements with a preceding rest in all patients at the initial visit; if AOBP readings remain <130 mm Hg in subsequent visits, measurements could be accepted, otherwise, if are higher, patients should be evaluated by out-of-office BP measurements.
More Related Videos
09:33A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...