Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Optimization of transmission and reconstruction parameters in angular displacement compounding using plane wave
Gijs A G M Hendriks1, Hendrik H G Hansen1, Chris L De Korte1
1Medical UltraSound Imaging Center (MUSIC), Department of Radiology and Nuclear Medicine, Radboud Institute for Health Sciences, Radboud University Medical Center, Nijmegen, The Netherlands.
None:
In ultrasound elastography, plane-wave acquisitions and angular displacement compounding (ADC) are often used and combined to allow high frame rates and to improve accuracy of lateral displacement estimates, respectively. This study investigates the performance of displacement and strain estimation for ADC as a function of; the main-to-grating-lobe-amplitude ratio which decreases as a function of steering angle; plane-wave acquisition and Delay-and-Sum (DaS)-related parameters; and grating-lobe filter cut-off frequency. Three experiments were conducted with a block phantom to test ADC performance for displacement fields of varying complexity: a lateral transducer shift, phantom rotation and phantom deformation. Experiments were repeated for four linear array transducers (pitch-to-lambda ratios between 0.6 and 1.4). Best ADC performance was found for steering angles that resulted in a theoretically derived main-to-grating-lobe-amplitude ratio of 1.7 dB for pure lateral translation and 6 dB for predominately lateral strain or rotation. Temporal filtering to reduce grating lobe signal or shifting of the receive aperture to receive angles below or above the optimal angle, as dictated by the main-to-grating-lobe-amplitude ratio, did not improve results. The accuracy of lateral displacement and strain estimates was improved by apodization in transmission and a dedicated F-number in DaS (0.75) allowing incidence angles within ± 33° in the active aperture. ADC with the optimized settings as found in this study improves the accuracy of displacements and strain estimates up to 80.7% compared to non-ADC. Compared to ADC settings described in current literature, our optimization improved the accuracy by 11.9% to 75.3% for lateral displacement and strain, and by 89.3% to 96.2% for rotation. The accuracy of ADC in rotation seemed to depend highly on plane-wave and DaS-related parameters which may explain the major improvement compared to settings in current literature. The overall improvement by optimized ADC was statistically significant compared to non-ADC (p = 0.003) and literature (p = 0.002).
More Related Videos
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
08:39Multiplexing Focused Ultrasound Stimulation with Fluorescence Microscopy
Published on: January 7, 2019
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Reconstruction of Signal using Interpolation
Relative Motion Analysis using Rotating Axes-Problem Solving
Here, in order to determine the magnitude of velocity and acceleration for point...
Linear Approximation in Time Domain
For a simple pendulum with a mass evenly distributed along its length and the center of mass located at half the pendulum's length,...
Transformation of Plane Strain
Under plane strain conditions, typical for members where one dimension significantly exceeds the others, deformations and resultant strains are...
Angular Velocity and Displacement