Hemoglobinometry: evaluation of a new method with a stable primary standard.

B R Wylie1, V A Lovric

  • 1Red Cross Blood Transfusion Service, Sydney.

Pathology
|April 1, 1988
PubMed
Summary

A new hemoglobinometry method using chlorohemin as a stable standard offers advantages over the cyanmethemoglobin technique. This method provides accurate hemoglobin measurements with a long-lasting, reliable standard.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Thanksgiving day for the gift of life.

The Medical journal of Australia·1997
Same author

Managing HIV. Part 8: controlling an epidemic. 8.4 Protecting the blood supply from HIV.

The Medical journal of Australia·1996
Same author

Effect of blood donation on the establishment of normal ranges of lymphocyte subsets.

Transfusion·1996
Same author

Multiple endoscopies in a Sydney blood donor found positive for hepatitis B and C antibodies.

The Medical journal of Australia·1996
Same author

Hepatitis B surface antigenaemia in a child after vaccination.

The Medical journal of Australia·1996
Same author

A positive hepatitis C enzyme immunoassay antibody test in a low risk population: what does it mean?

The Medical journal of Australia·1995

Area of Science:

  • Clinical Chemistry
  • Analytical Chemistry

Background:

  • The cyanmethemoglobin method is a standard for hemoglobin measurement but has limitations.
  • A stable primary standard for hemoglobinometry is needed.

Purpose of the Study:

  • To evaluate a new hemoglobinometry method using purified chlorohemin as a stable standard.
  • To compare the new method with the cyanmethemoglobin technique.

Main Methods:

  • Absorption spectra of blood samples and chlorohemin were analyzed in an alkaline-detergent solution.
  • Stability of the reagent mixture and chlorohemin standard was assessed under various storage conditions.
  • The new method was correlated with the cyanmethemoglobin technique.

Main Results:

More Related Videos

A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering
05:18

A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering

Published on: December 7, 2016

A Point-of-Care Method with Integrated Decision Support Tool to Estimate Anemia at Population Level
05:35

A Point-of-Care Method with Integrated Decision Support Tool to Estimate Anemia at Population Level

Published on: January 19, 2024

Related Experiment Videos

Last Updated: Jul 22, 2026

Staphylococcus aureus Growth using Human Hemoglobin as an Iron Source
06:37

Staphylococcus aureus Growth using Human Hemoglobin as an Iron Source

Published on: February 7, 2013

A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering
05:18

A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering

Published on: December 7, 2016

A Point-of-Care Method with Integrated Decision Support Tool to Estimate Anemia at Population Level
05:35

A Point-of-Care Method with Integrated Decision Support Tool to Estimate Anemia at Population Level

Published on: January 19, 2024

  • Chlorohemin and blood samples exhibited identical absorption spectra with a peak at 575 nm.
  • The reagent and chlorohemin standard remained stable for at least 250 days.
  • The new method showed close correlation with the cyanmethemoglobin technique, offering advantages such as a stable standard and improved conversion.
  • Conclusions:

    • The newly described hemoglobinometry method using chlorohemin offers significant advantages over the cyanmethemoglobin method.
    • This method provides a stable primary standard with a long shelf-life for accurate hemoglobin measurements.