Disadvantages of Automated Respiratory Gas Exchange Analyzers
Juan José Ramos-Álvarez1, Irma Lorenzo-Capellá2, Francisco Javier Calderón-Montero3
1School of Sports Medicine, Madrid Complutense University, Madrid, Spain.
Abstract:
The use of automated gas exchange analysis systems in exercise studies is common throughout the industrialized world and are frequently used in sports medicine laboratories for the measurement of maximal oxygen uptake (VO2max), as an integrative parameter that allows the physical condition to be assessed, in spite of its limitations. Actually, the fundamental principles behind the measurement of respiratory gas exchange (RGE) have not changed for a century. It was used a manual Douglas bag method together with separate chemical analyses. The need for faster and more efficient techniques, has conditioned the traditional procedures and determined the emergence of automated systems. However, the validity and reliability of all these different systems is not well known. The common features associates with these systems, also have disadvantages that must be evaluated at the time of the acquisition of an automated equipment: (1) regular quality control checks, which entails other added economic costs, (2) the validity and reliability of the results, which it is necessary to verify, and (3) the user does not know the equations that determine the values of oxygen consumption and carbon dioxide production. This work aims to clarify the disadvantages of these automated systems. At maximum intensities, the variation of VO2max or VO2peak can be very significant in athletes and even more relevant in sick people undergoing a training program. Therefore, considerable care is needed when comparing RGE data with automated systems.
New And Noteworthy:
Actually, stress tests are more conveniently performed with automated systems. It is necessary to examine the validity and reliability of automated respiratory gas exchange systems. The algorithms incorporated in the software, apart from being a "mystery," show differences with respect to the data provided.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
External and Internal Respiration
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...