Reliability of different electromyographic normalization methods for pelvic floor muscles assessment
Vanessa S Pereira-Baldon1,2, Ana B de Oliveira1, Juliana F Padilha1
1Physical Therapy Department, Universidade Federal de São Carlos, São Carlos, São Paulo, Brazil.
Reliable pelvic floor muscle (PFM) electromyography (EMG) normalization is crucial. Peak PFM maximal voluntary contraction (MVC) or abdominal contraction RMS values offer excellent test-retest reliability for PFM-EMG studies in young women.
Area of Science:
- Physiology
- Biomechanics
- Medical Engineering
Background:
- Electromyography (EMG) normalization is essential for accurate physiological signal interpretation.
- Pelvic floor muscle (PFM) dysfunction affects a significant portion of the population, necessitating reliable assessment methods.
- Standardized EMG normalization protocols are vital for reproducible research and clinical evaluation of PFM function.
Purpose of the Study:
- To assess and compare the test-retest reliability of various normalization methods for PFM electromyography (EMG).
- To identify the most reliable normalization techniques for PFM-EMG data in healthy nulliparous women.
- To provide evidence-based recommendations for PFM-EMG data processing in research and clinical settings.
Main Methods:
- Thirty healthy nulliparous women participated in two EMG testing sessions separated by seven days.
- Vaginal probe in situ recorded PFM EMG signals during maximal voluntary contraction (MVC) and three daily activities (coughing, Valsalva, abdominal contraction).
- Intraclass correlation coefficients (ICC), SEM, %SEM, and MDC were calculated to evaluate reliability.
Main Results:
- Test-retest reliability (ICC) for normalization methods ranged from 0.61 to 0.95.
- Highest reliability was observed for mean RMS of abdominal contraction and peak RMS of PFM-MVC.
- Normalization using PFM-MVC RMS yielded the lowest SEM and MDC values, indicating superior precision.
Conclusions:
- Normalization of PFM-EMG using peak RMS of PFM-MVC or mean/peak RMS of abdominal contraction demonstrates excellent reliability.
- These recommended normalization methods are suitable for studies evaluating PFM function in young women.
- Establishing reliable EMG normalization enhances the validity and comparability of PFM research.