Safety and Efficacy of a Novel Pneumatically Driven Extracorporeal Membrane Oxygenation Device
Christian Karagiannidis1, Thilo Joost2, Stephan Strassmann1
1Department of Pneumology and Critical Care Medicine, Cologne-Merheim Hospital, ARDS, Cologne, Germany; ECMO Centre, Kliniken der Stadt Köln gGmbH, Witten/Herdecke University Hospital, Cologne, Germany.
Background:
Extracorporeal membrane oxygenation (ECMO) is rapidly becoming a mainstream technology for lung or heart/lung support. Current ECMO devices mostly consist of a power-driven centrifugal pump and a dedicated oxygenator. We studied the safety and efficacy of a novel, fully pneumatically driven ECMO device, which could be used in both venovenous or venoarterial mode in an animal model.
Methods:
Six healthy, awake sheep were treated with the Mobybox ECMO device (Hemovent, Aachen, Germany) over a 7-day period in a venovenous mode. Gas exchange, coagulation parameters, and safety were assessed.
Results:
Using a blood flow rate of 2 L/min and a low sweep gas flow rate of 0.3 L/min, the PCO2 ranged from 38 to 44 mm Hg pre oxygenator and dropped to 32 to 36 mm Hg post oxygenator, whereas the PaO2 post oxygenator increased to 600 mm Hg. Higher levels of sweep gas flow resulted in cessation of spontaneous breathing in some animals, consistent with high-efficiency carbon dioxide removal; thus, the sweep gas flow rate was maintained at a low level. Platelets dropped from 177 ± 53/μL to 107 ± 28/μL on day 2, while returning to baseline by day 7 (180 ± 51/μL). Plasma-free hemoglobin remained low (2-9 mg/dL), whereas fibrinogen slightly increased, and then remained stable throughout the period. Neither the pump nor the oxygenator showed any visible clotting after 7 days.
Conclusions:
The pneumatically driven ECMO device provided excellent safety and physiologic efficacy in a 7-day sheep experiment without visible clotting, hemolysis, or sustained reductions in fibrinogen or platelets.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: