Are implantable cardiac monitors reliable tools for cardiac arrhythmias detection? An intra-patient comparison with
Antonio Bisignani1, Silvana De Bonis2, Luigi Mancuso3
1Department of Cardiovascular and Thoracic Sciences, Fondazione Policlinico Universitario A. Gemelli IRCCS, Catholic University, Rome, Italy.
Introduction:
Implantable cardiac monitor (ICM) is an established tool for the management of unexplained syncope and atrial fibrillation (AF) even if its accuracy of arrhythmia detection may be suboptimal. The aim of this study was to perform an intra-patient comparison of the diagnostic capability of ICM with a dual-chamber PM as a gold standard.
Methods:
We included 19 patients with a previously implanted ICM (BioMonitor 2 Biotronik, Berlin, Germany), who received a dual-chamber PM for standard indications. ICM-detected arrhythmic events in a 6-month follow-up were compared with those detected by the PM and classified by visual inspection of intracardiac electrograms.
Results:
During follow up, ICMs generated 15 false asystole and 39 false bradycardia detections in 5 patients (26.3%) due to recurrent premature ventricular contractions. A total of 34 true high ventricular rate (HVR) episodes were detected by the PM. Of them, 30 (88%) events were also recorded by the ICM, which further detected 14 false HVR snapshots, leading to a sensitivity and positive predictive value of 88% and 68%, respectively. In addition, PM identified 234 true AF episodes. Of them, 225 (96%) events were also detected by the ICM, while 8 (42%) ICMs stored 50 AF episodes classified as false positives. The ICM sensitivity for AF was 96% with a positive predictive value of 82%.
Conclusion:
Our intra-patient comparison with permanent PM confirmed that ICM is an effective tool for cardiac arrhythmias detection. ICM algorithms for AF and HVR detection were highly sensitive with an acceptable rate of false positive episodes.
More Related Videos
08:56Sterile Pericarditis in Aachener Minipigs As a Model for Atrial Myopathy and Atrial Fibrillation
Published on: September 24, 2021
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias