Background:
ST-segment deviation post-electrical cardioversion is a common finding amongst a significant number of patients. However, the mechanism by which this phenomenon occurs and its clinical implications are not fully understood.
Case Summary:
Four patients presented to our department with complaints of palpitations. They were found to have atrial fibrillation and were cardioverted using a synchronized direct current shock at 200 J. However, their telemetry immediately after the shock showed transiently an ST-segment elevation resembling Brugada Type 1 pattern. All telemetries normalized within 6 s from cardioversion.
Discussion:
Through this report, we aim to raise mild concern about a possible safety issue related to synchronized electrical cardioversion with electrodes positioned in an anteroposterior fashion. Also, we try to give a pathophysiological explanation to this finding on the base of the knowledge accumulated through the years on Brugada syndrome.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
05:12Transesophageal Atrial Burst Pacing for Atrial Fibrillation Induction in Rats
Published on: February 14, 2022
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Dysrhythmias II: Classification of Tachyarrhythmias