Related Experiment Video
Updated: Dec 27, 2025

Synthetic, Multi-Layer, Self-Oscillating Vocal Fold Model Fabrication
Published on: December 2, 2011
Effects of thyroplasty implant stiffness on glottal shape and voice acoustics
Brian H Cameron1, Zhaoyan Zhang2, Dinesh K Chhetri2
1Keck School of Medicine of USC Los Angeles, California.
Objectives:
Vocal fold (VF) stiffness and geometry are determinant variables in voice production. Type 1 medialization thyroplasty (MT), the primary surgical treatment for glottic insufficiency, changes both of these variables. Understanding the cause and effect relationship between these variables and acoustic output might improve voice outcomes after MT. In this study, the effects of thyroplasty implants with variable stiffness on glottal shape and acoustics were investigated.
Methods:
In an ex vivo human larynx phonation model, bilateral MT with implants of four stiffness levels (1386, 21.6, 9.3, and 5.5 kPa) were performed. Resulting acoustics and aerodynamics were measured across multiple airflow levels. A vertical partial hemilaryngectomy was performed and stereoscopic images of the VF medial surface taken to reconstruct its three-dimensional (3D) surface contour. The results were compared across implants.
Results:
The effects of implant stiffness on acoustics varied by airflow. Softer implants resulted in improved acoustics, as measured by cepstral peak prominence (CPP), at lower airflow levels compared to stiffer implants but this relationship reversed at high airflow levels. Stiffer implants generally required less airflow to generate a given subglottal pressure. Stiffer implants resulted in greater medialized surface area and maximal medialization, but all implants had similar effects on overall VF medial surface contour.
Conclusion:
Softer implants result in less medialization but better acoustics at low airflow rates. Stiffer implants provide better acoustics and more stable pressure-flow relationships at higher airflow rates. This highlights a potential role for patient-specific customized thyroplasty implants of various stiffness levels.
Level Of Evidence:
NA.
More Related Videos
06:24Author Spotlight: Advancements in the Fabrication of Synthetic Vocal Fold Models for Phonetic and Robotic Applications
Published on: January 5, 2024
07:22Minimally Invasive Murine Laryngoscopy for Close-Up Imaging of Laryngeal Motion During Breathing and Swallowing
Published on: December 1, 2023
Related Concept Videos
Larynx
Anatomy of the Larynx
The larynx consists of various components, including cartilage, muscles, and vocal cords. Its structure includes three large unpaired cartilages—the thyroid, cricoid, and epiglottis—and three smaller paired cartilages—the arytenoids,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...