Published on: February 9, 2012
An analytical solution to the dispersion-by-inversion problem in magnetic resonance elastography.
Joaquin Mura1, Felix Schrank2, Ingolf Sack2
1Department of Mechanical Engineering, Universidad Técnica Federico Santa María, Santiago, Chile.
Finite difference operators in Magnetic Resonance Elastography (MRE) cause biased stiffness estimates. This study analytically and experimentally shows how to minimize these artifacts by optimizing parameters based on noise and wave properties.
Area of Science:
- Biomedical Engineering
- Medical Imaging
- Rheology
Background:
- Magnetic Resonance Elastography (MRE) quantifies soft tissue stiffness using shear vibrations.
- Current MRE methods often employ inversion-based reconstructions with finite difference operators (FDOs).
- FDOs can introduce biased frequency dispersion in stiffness estimates.
Purpose of the Study:
- To analytically, numerically, and experimentally investigate the impact of FDOs on MRE stiffness estimates.
- To identify the sources of bias in FDO-based MRE reconstructions.
- To provide a method for minimizing inversion artifacts in MRE postprocessing.
Main Methods:
- Analysis of spatial harmonic response to shear vibrations in MRE.
- Application of first- and second-order finite difference operators (FDOs).
- Analytical derivation and experimental validation of FDO dispersion functions.
Main Results:
- FDOs lead to underestimation at low frequencies (high spatial support) and overestimation at high frequencies (low spatial support) due to noise and discretization.
- Second-order FDOs are more sensitive to noise than first-order FDOs.
- Optimized parameter A (wavelength/pixel size/stencil pixels) minimizes FDO artifacts, with specific SNR-dependent ranges provided for first- and second-FDOs.
Conclusions:
- This study offers an analytical solution to the long-standing problem of FDO-induced artifacts in MRE.
- The findings contribute to reducing inversion artifacts, improving the accuracy of MRE stiffness measurements.
- Optimizing FDO parameters based on SNR and wave properties is crucial for reliable MRE postprocessing.